8-5. Function list

Class Item

Param-

eter

Description

Š

0

Auto power-off disabled

poff

Auto power-off

1

Auto power-off enabled

Automatically
power off

0

Fast / sensitive

1

Š

2

3

Cond

Response

4

Slow / stable

Software filtering

0

Stable when within ± 0.5d/0.5s

Š

1

Stable when within ± 1d/0.5s

st-b

Stability band
width

2

Stable when within ± 2d/0.5s

Conditions to
turn on the
STABLE indicator

0

Disabled

trc

Zero tracking

Š

1

Enabled

Tracking zero shift

Š

0

Point (.)

pnt

Decimal point

1

Comma (,)

Decimal separator

Š

0

Comparator disabled

1

Compares all data

2

Compares all stable data

3

Compares plus data > +4d

4

Compare stable plus data > +4d

5

Compares data > +4d or < -4d

Cp

Comparator mode

6

Compares stable data > +4d or < -4d

Conditions to
compare.
d = the minimum
display division

0

Command and stream modes

Š

1

Command and PRINT key

2

Command, PRINT key and auto-print A

3

Command, PRINT key and auto-print B

prt

Data output
mode

4

Command mode only

Auto-print A:

+ data

Auto-print B:

+/- data

Š

0

No pause (general equipment)

pUse

Data output pause

1

1.6 seconds (for AD-8121)

Interval between
continuous data

Š

0

No output

1

AD-8121 format

(

*

)

info

GLP output

2

General format

GLP
output format

Š

0

2400 bps

1

4800 bps

2

9600 bps

bps

Baud rate

3

1200 bps

Š

0

7 bits, even parity

1

7 bits, odd parity

func

btpr

Data and parity

2

8 bits, non parity

Š

Factory setting

(

*

)

When the AD-8121 format is selected, the interval between data is 1.6 seconds
regardless of the setting “pUse”.