Cc 0 0.0 cc 0 – A&D Ej Series Compact Balance EJ-2000 User Manual

Page 35

Advertising
background image

33


2. CC 0 is displayed.

3. If necessary to change the weight value to check, press the SAMPLE key and

change the weight value using the following keys.

SAMPLE key

To select the digit blinking to change.

RE-ZERO key To set the value of the digit selected.

PRINT key

To store the value and return to step 2.


4. At step 2, press the PRINT key. The zero point

is weighed and the weighed value is displayed for a
few seconds.








5. Place the displayed weight on the pan and press

the

PRINT key to weigh it. The weighed value is

displayed for a few seconds.















6. end appears.


7. glp is displayed and calibration test report is

output.



8. end appears again. Remove the weight and

press the MODE key to return to the weighing
mode.



CC 0

0.0

CC 0

The weighed data

is shown with “g”.

With nothing

on the pan

2000.0

2000.0

2000.0

Displayed

weight

end

end

The weighed data

is shown with “g”.

Returns to

the weighing mode

glp

GLP output

Advertising