7. CALIBRATION

This function adjusts the balance for accurate weighing.
Perform a calibration in the following cases.

‰

When the balance is first used.

‰

When the balance has been moved.

‰

For regular calibration.









7-1. Calibration using a weight

‰

Prepare the calibration weight (sold separately) in advance.

1. Warm up the balance for at least half an hour with

nothing on the pan.

2. Press and hold the calibration (CAL) switch until

Cal appears, and release the switch.


3. The balance displays Cal 0 .


To change the calibration weight value, proceed to
step 4.

To use the calibration weight value in the balance
memory, proceed to step 5.

4. Press the SAMPLE key. The display shows the

calibration weight value in “gram” that is stored in the
balance. Use the following keys to change the value.

SAMPLE key To select the digit blinking to

change.

RE-ZERO key To set the value of the digit

selected.

PRINT key

To store the value and return to
step 3.

UNITS key

To cancel the value and return to
step 3.



Cal

Cal 0

06000.0

Cal 0

Set the weight using

the relevant keys.

Release the CAL switch.

Calibration (CAL) switch

Calibration (CAL) switch cover

Press and hold

the CAL switch.