10. ID NUMBER AND GLP

‰

The ID number is used to identify the balance when Good Laboratory Practice

(GLP) is used.

‰

The ID number is held in memory even if power to the balance is switched off.

‰

The following GLP data is transmitted to a printer or a computer using the optional

RS-232C interface.

• The result of calibration (“Calibration report”)
• The result of calibration test (“Calibration test report”)
• The “Title block” and “End block” for GLP data

‰

The GLP output format includes the balance manufacturer name, model number,

serial number, ID number and space for signature.

‰

When used with the AD-8121B, the date and time can be printed (GLP output

format info = 1).

10-1. Setting the ID number

1. Press and hold the SAMPLE key to display

func .


2. Press the SAMPLE key several times to display

id .


3. Press the PRINT key. Enter the ID number using

the following keys.

SAMPLE key To select the digit blinking to

change.

RE-ZERO key To set the character of the digit

selected. See the table below for
the “display character set”.

PRINT key

To store the value and proceed to
the next step.

UNITS key

To cancel the value and proceed
to the next step.


4. When the above operation has completed, func

appears after end .


5. Press the UNITS key to return to the weighing

mode.

Display character set

0 1 2 3 4 5 6 7 8 9 - A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
0 1 2 3 4 5 6 7 8 9 - _ a b C d e f g H i j k l m n o p q r s t U v w x y z

“ ” : Space

Press

several times

func

id

000000

end

id

func

Set using the

relevant keys

Returns to

the weighing mode