Data format

S T

,

+

0

0

0

0

0

.

0

0

g C

R

L

F

Header

Data

Unit Terminator



‰

There are four types of headers:

ST : Stable weighing data (including % data)

QT : Stable counting data

US : Unstable weighing data (including count and %)

OL : Out of weighing range (Over)


‰

The data is normally 9 digits including a decimal point and a sign.


‰

There are 14 types of units:

g : Weighing data “gram”

P C : Counting data “pcs”

% : Percentage data “%”

o z : Weighing data “decimal ounce”

l b : Weighing data “decimal pound”

o z t : Weighing data “troy ounce”

c t : Weighing data “carat”

m o m : Weighing data “momme”

d w t : Weighing data “penny weight”

G N : Weighing data “grain”

N : Force data “Newton”

t l : Weighing data “tael”

t : Weighing data “tola”

DS

: Calculated density (specific gravity) value


‰

The terminator is always C

R

L

F

.


‰

Example of output data:

Weighing data “gram”

S T ,

+ 0 0 1 2 3 4

.

5

g C

R

L

F

Counting data

Q T ,

+ 0 0 0 1 2 3 4 5

P C C

R

L

F

Percentage data

S T ,

+ 0 0 0 1 2 3

. 4

% C

R

L

F

Out of range “gram” (+)

O L

,

+ 9 9 9 9 9 9

. 9

g C

R

L

F

Out of range “pcs” (-)

O L

,

- 9 9 9 9 9 9 9 9

P C C

R

L

F

Data output mode

‰

Stream mode

Set the function “prt 0”.
The balance outputs the current display data. The data-update rate is approximately
10 times per second. This rate is the same as the display-update.
The balance does not output data while it is in the setting mode.

Separator

(“ ” shows a space.)