Unit – A&D Ej Series Compact Balance EJ-2000 User Manual

Page 15

Advertising
background image

13

Entering the upper and lower limit values

6. With Cp Hi displayed, press the PRINT key.

Enter the upper limit value using the following keys.

SAMPLE key To select the digit blinking to

change.

RE-ZERO key To set the value of the digit

selected. Hold down the key to
switch the sign “+” and “-“.

(“

N” designates a negative

value.)

PRINT key

To store the value and proceed to
the next step.

UNITS key

To cancel the value and proceed
to the next step.






7. With Cp lo displayed, press the PRINT key.

Enter the lower limit value using the following keys.

SAMPLE key To select the digit blinking to

change.

RE-ZERO key To set the value of the digit

selected. Hold down the key to
switch the sign “+” and “-“ (see
step 6).

PRINT key

To store the value and proceed to
the next step.

UNITS key

To cancel the value and proceed
to the next step.



8. Press the PRINT key. Unit appears after

end .






9. Press the UNITS key to return to the weighing

mode.




Cp lo

end

000000

Set using the

relevant keys

001230

LO

LO

Unit

Returns to

the weighing mode

To store

Cp Hi

000000

Set using the

relevant keys

HI

001234

HI

To store

Advertising