7-2. Gravity acceleration correction

When the balance is first used or has been moved to a different place, it should be
calibrated using a calibration weight.
But if the calibration weight cannot be prepared, the gravity acceleration correction will
compensate the balance. Change the gravity acceleration value of the balance to the
value of the area where the balance will be used. See the gravity acceleration map
appended to the end of this manual.

‰

Note

Gravity acceleration correction is not required when the balance is

calibrated using a calibration weight at the place where the balance
is to be used.



1. Press and hold the calibration (CAL) switch until

Cal appears, and release the switch.


2. The balance displays Cal 0 .


3. Press the RE-ZERO key.

The display shows the gravity acceleration value
stored in the balance.
Use the following keys to change the value.

SAMPLE key To select the digit blinking to

change.

RE-ZERO key To set the value of the digit

selected.

PRINT key

To store the value and return to
step 2.

UNITS key

To cancel the value and return to
step 2.


4. After setting the value, press the PRINT key.

Cal 0 is displayed again.


5. If it is necessary to calibrate the balance using a

calibration weight, go to step 4 of 7-1.
To finish the setting, press UNITS key.


6. end appears and the balance returns to the

weighing mode.








Cal

Cal 0

9.7985

Cal 0

Returns to

the weighing mode

To store

Release the CAL switch.

Set the value using

the relevant keys.

Press and hold

the CAL switch.