15

5. At step 3, pressing the PRINT key to weigh the

zero-point value. Do not touch the pan during
weighing.













6. Place the calibration weight with the same value as

displayed on the pan. Press the PRINT key to
weigh it. Do not touch the pan during weighing.









7. end appears.

Remove the weight from the pan.
When the GLP output (function setting “info 1” or
“info 2”) is selected, glp is displayed. The
calibration report is output and end appears
again.
Press the CAL switch or UNITS key to return to
the weighing mode.








‰

Note

The value set in step 4 is stored in memory even after the power is

switched off.

If the balance is to be moved to other places, set the gravity

acceleration value of the area where the calibration using a weight is
to be done, and calibrate the balance according to the procedure
above. See the next section to set the value.



5000.0

0

end

000

With nothing

on the pan

To store

To store

Calibration

weight

Cal 0

5000.0

glp

GLP output

0

end

000

Returns to

the weighing mode