APPENDIX A

AALBORG

7 VORTEX METER EEPROM Variables Rev.P002 [07/08/2010]

Note: indexes 0-19 are write protected (manufacture and calibration specific data)

INDEX

NAME

DATA
TYPE

NOTES

0

BlankEEPROM[10]

char[10] Do not modify. Table Revision [PROTECTED]

1

SerialNumber[20]

char[20] Serial Number [PROTECTED]

2

ModelNumber[20]

char[20] Model Number [PROTECTED]

3

SoftwareVer[10]

char[10] Firmware Version [PROTECTED]

4

ManufReservedF1

float

Manufacture Specific float data [PROTECTED]

5

ManufReservedF2

float

Manufacture Specific float data [PROTECTED]

6

MeterCalFactorM

float

Original Manuf. Assigned Meter Cal Factor
[PROTECTED]

7

MeterSize

float

Original Manuf. assigned size of the meter [inch]
[PROTECTED]

8

ReservedText[12]

char[12]

Reserved for Manufacture Specific Text Info
[PROTECTED]

9

ManufReservedF3

float

Manufacture Specific float data [PROTECTED]

10

ManufReservedF4

float

Manufacture Specific float data [PROTECTED]

11

ManufReservedF5

float

Manufacture Specific float data [PROTECTED]

12

ManufReservedF6

float

Manufacture Specific float data [PROTECTED]

13

ManufReservedUI1

uint

Manufacture Specific uint data [PROTECTED]

14

ManufReservedUI2

uint

Manufacture Specific uint data [PROTECTED]

15

ManufReservedUI3

uint

Manufacture Specific uint data [PROTECTED]

16

ManufReservedUI4

uint

Manufacture Specific uint data [PROTECTED]

17

ManufReservedSI1

int

Manufacture Specific int data [PROTECTED]

18

ManufReservedSI2

int

Manufacture Specific int data [PROTECTED]

19

ManufReservedSI3

int

Manufacture Specific int data [PROTECTED]

20

TimeSinceCalHr

float

Time elapsed since last calibration in hours

21

ProtectionCode

uint

Program Parameters Protection Code [0-255]

22

DPTreshLevel

uint

Auto DP attenuator threshold level (+/- counts from target)

23

BackLightMode

uint

Reserved for future implementation**

24

LCD_Diagnostic

uint

LCD Diagnostic Mode: [0, 1]

25

Address485

char[4] Two hexadecimal characters address for RS485 only [01-FF]

26

LCD_Mode

uint

Main Process screen mode: 0-Static,1-Dynamic

27

LCD_Static_Mode

uint

Type of the static LCD screen: 0-12

28

LCD_AUTO_Mode_Mask

uint

Mask of the AUTO LCD screen: hex 0x01 - 0xFF

29

LCD_Cycle_Tim

uint

Time in seconds for each var. to be displayed in AUTO mode

30

Out4_20mA_Mode1

uint

Process Variable assignment to 4-20mA output channel #1: [0-5]

31

Out4_20mA_Mode2

uint

Process Variable assignment to 4-20mA output channel #2: [0-5]

Advertising