Aalborg Vortex Insertion User Manual
Page 88
83
COMMAND
NAME
DESCRIPTION
NO.
COMMAND SYNT
A
X
COMMAND
ARGUMENT 1
ARGUMENT 2
ARGUMENT 3
ARGUMENT 4
RESPONSE
Te
mperature
Alarms**
Sets / reads the status of the
temperature alarms.
Note: High and Low limits have
to be entered in the °C. High
alarm value has to be more
than Low alarm value.
Alarm conditions:
Temp. > High Limit = H
Temp. < Low Limit = L
Low < T
emp. < High = N
9
TA
H
(high temp. limit)
<V
alue>
(-20.1-260 °C)
TA
H:<V
alue [°C]>
L
(low temp. limit)
<V
alue>
(-20.1-260 °C)
TA
L:<V
alue [°C]>
A
(action delay in
seconds)
<V
alue> (0-3600
sec.)
TAA:<V
alue [Sec]>
E
(enable alarm)
TA
:E
D
(disable alarm)*
TA
:D
R (read current
status)
TA
R:N (no alarm)
TA
R:H (high alarm)
TA
R: L (low alarm)
S (Read current
settings)
TAS:M,L,H,D,B where:
M – mode (E/D)
L – Low settings (°C)
H – High settings (°C)
D – Action Delay (sec)
B – Latch mode (0-1)
B Block (Latch)
mode
<V
alue>
(0-disabled*)
(1-enabl’d )
TAB:<V
alue> where:
V
alue = 0 – 1