Aalborg Vortex Insertion User Manual
Page 99
94
COMMAND
NAME
DESCRIPTION
NO.
COMMAND SYNT
A
X
COMMAND
ARGUMENT 1
ARGUMENT 2
ARGUMENT 3
ARGUMENT 4
RESPONSE
FFT DSP
Settings
Sets/Reads FFT DSP
related variables.
FFT Sample/Second:
0 – Automatic
1 – 8000 samples/sec
2 – 4000 samples/sec
3 – 2000 samples/sec
4 – 1000 samples/sec
5 – 500 samples/sec
21
FT
M
FFT Algorithm Mode:
E- Enable
D- Disable
<New V
alue>
E or D
FTM:<value>
Example: FTM:E
No Argument
(Ret. Current settings)
FTM:<value>
Example: FTM:E
T
FFT T
rigger Level:
0 – Auto
[1-4095] - Manual
<new value>
[0-4095]
FTT
:<value>
Example: FTT
:60
No Argument
(Ret. Current settings)
FTT
:<value>
Example: FTT
:60
S
FFT Sample/Second
(manufacture only
, do
not change*)
<new value>
[0-5]
FTS:<value>
Example: FTS:1
No Argument
(Ret. Current settings)
FTS:<value>,<SF>
Example: FTS:1,1000
D
FFT Damping
(manufacture only
, do
not change*)
<New V
alue>
E or D
FTD:<value>
Example: FTF:D
No Argument
(Ret. Current settings)
FTD:<value>
Example: FTF:D
N
FFT Minimum Noise
Threshold for Auto
(manufacture only
, do
not change*)
<new value>
[0-5000]
FTN:<value>
Example: FTN:500
No Argument
(Ret. Current settings)
FTN:<value>
Example: FTN:500
A
Attenuation Mode and
Ta
rget Level
<value>
A or M
<value>
[0-4095]
FT
A:<value>,<value>
Example: FT
A:M,1000
No Argument
(Ret. Current settings)
FT
A:<value>,<value>
Example: FT
A:M,1000