NOTE: 3 1/2 digit LCD may be replaced with either a 3 1/2 or a

4 1/2 digit backlit LCD - available as options.

ACCURACY

Analog Models: + 2% of Full Scale

Digital Models: + 1 % of Full Scale

SENSITIVITY

0.05% of span

STABILITY

0.05% of span

REPEATABILITY

0.1% of span

CALIBRATION CHECK

Built in full scale value

RECORDER OUTPUT

0-10 VDC @ 5mA max. (linear); standard on all models

OUTPUT IMPEDANCE

100Ω ±5%

SENSOR INPUT 1 (optional 3 sensor input available)
CONTROL FUNCTION

Models 757II, 758II & 752II, 753II:

Single set point alarm/control continuously adjustable

0-100% of span

Second set point alarm/control (Optional).

Continuously adjustable - 0-100% of span

Hysteresis

Adjustable from 0.3-3% of full scale

Indicators

“HIGH” (red) and “LOW” (green) set point LEDs - reversible.

Relay Contact Rating

SPDT 10 amp @ 250 VAC, 30 VDC. Relay operates

increasing or decreasing reading (user selectable).

Solid State Output

24VDC unregulated, 30mA Maximum.

Powers optional PA - Piezo Electric Alarm, RA™ - Remote

Alarm, or ?

Second Alarm/control Module, with above specifications, opt. on

Models 757II, 758II & 752II, 753II

POWER SPECIFICATIONS

115/230 VAC ±15%, 50/60 Hz (User selectable)

overvoltage category II

100 mA Maximum Current

Double Insulated (with circuit board ground for OEM

operation)

Fuse - 100mA Slow Blow (T.10A) for both 115 & 230 VAC

(V~) User replaceable

Humidity - 20-90% non-condensing

Max. Altitude -

12,000 meters/40,000 ft non-operating

3000 meters/10,000 ft operating

Pollution degree 2

24 VAC or 24 VDC Option available on Special Order

Overvoltage category II (24 VAC)

250 mA Maximum Current

Double Insulated (with circuit board ground for OEM

operation)

Fuse - 250mA Slow Blow (T.25A) User replaceable

AMBIENT TEMPERATURE RANGE

0-60ºC/32-140ºF

DIMENSIONS

152H x 122W x 96 D mm / 6.0 H x 4.8 W x 3.8 D in.

HOUSING CONSTRUCTION

Fully gasketed heavy-duty ABS for splash-proof and corrosion

resistance.

Rated IP65/NEMA 4X

Double Insulated (with circuit board ground for OEM operation)

WEIGHT

750 Series II: average 0.9 kg/2 Ibs.

C.

OPTIONAL FEATURES

See sections VI & VII for complete selection and details.

-SC

Second Alarm/control Module (M/c only)

-4A

4-20 mA Isolated output (digital models only)

-PA

70 db Piezo Electric Alarm (digital M/c only)

-PAT

Piezo Alarm & Timer Module (digital M/c only)

-45BL

4 1/2 Digit Backlit Liquid Crystal Display (LCD)

-35BL

3 1/2 Digit Backlit Liquid Crystal Display (LCD)

-TP

Temperature Module 0-200ºC, requires -TP sensor

(digital Monitor/controllers only)

-TH

Alarm/control Interface Harness (req. SC & TP)

(digital only)

-PC

115 VAC Power cord (8 ft./2,44 meters) with USA

plug and strain relief - NOT for use with 230 VAC.

-24VA

24 VAC Isolated Power Supply (special order)

-24VD

24 VDC Isolated Power Supply (special order)

-PTS

Panel mounted Full Scale Test switch (special order)

D.

ACCESSORIES (ordered separately)

See sections VI & VII for complete selection and details.

ASI

AQUASWITCH I, auto RO/DI switching system (req. M/c)

SCO

Second Alarm/control Module (Monitor/controllers only)

4AO

4-20 mA Self-Powered Isolated Output Module

(digital only)

3SE

3 Sensor Input Module with Enclosure.

3SO

3 Sensor Input Module with 1 meter/3 foot cable*.

3RE

3 Range Switch with Enclosure†

3RO

3 Range Switch with 1 meter/3 foot cable†*.

3SRO

Above 3SO & 3RO combined with enclosure†.

3SRE

Above 3SO & 3RO combined†*.

PAO

70 db Piezo Electric Alarm only (753II/758II)

PATO

Piezo Alarm & Timer Module (753II/758II)

45BLO

4 1/2 Digit Backlit Liquid Crystal Display (LCD)

35BLO

3 1/2 Digit Backlit Liquid Crystal Display (LCD)

TPO

Temperature Module 0-200ºC - requires -TP

sensor (digital models only)

THO

TP/TPO Alarm/control Harness only

(requires SCO & TPO)

PCO

Power Cord 115 VAC (8 ft./2,44 meters) with USA

plug and strain relief - NOT for use with 230 VAC.

SMP50 Surface Mounting Plate

RA

Remote Alarm™ - RA (Monitor/controllers only)

VR

Power Supply, 24 VAC, 20VA. (115VAC, indoor use)

024-1

1 in. NPT 24 VAC Solenoid Valve

CS-11

20Ω Resistivity Calibration Module (NIST Traceable)

CS-14

2Ω Resistivity Calibration Module (NIST Traceable)

CS-17

200Ω Resistivity Calibration Module (NIST Traceable)

†Add C for Conductivity or R for Resistivity to part number