3. CHANGING ANALOG METER SCALE (DIAL)

After changing ranges on an analog model 750II it may be nec-

essary to change the meter scale (dial). The following steps will

guide you through the change. Please read caution carefully.

Exercise extreme caution while working on meter. Meter

movement, and/or pointer (needle) may be damaged beyond

repair. It is recommended only qualified personnel change

scales.

DO NOT allow dust to enter the case while it is open.

Failure to follow these instructions may cause irreparable

damage, and will VOID the warranty.

Briefly -

The meter is removed from front panel (except OEM models).

The cover is removed.

The hex screws retaining the scale are removed, and scale

is slid out the top.

New scale is slid in place, and screws reinstalled.

Cover aligned, installed and zero set.

Physical

For OEM models skip to #4.

1. Using a wrench, nut driver or pliers remove the two

(2) large nuts retaining the circuit board wiring harness

connectors. Red wire is PLUS (+),

Brown wire is MINUS (-). See figure II.G.4.

2. Using a standard slot screwdriver remove the four (4)

screws, and the backing plate holding the meter to the

front panel as shown in figure II.G.4.

3. Remove meter. Do not lose the gasket between the

meter and the front panel it must be reinstalled upon

reassembly. See figure II.G.5. Align gasket to front

panel so as the gasket does not show when meter is

assembled to front panel.

4. Using a wide blade standard slot screwdriver, starting

at a bottom corner, CAREFULLY pry the clear plastic

cover loose from the movement case. See figure II.G.6.

5. Using a 1/8” (3.2mm) hexdriver, remove the two (2)

small screws holding the scale (dial). See figure II.G.7.

6. To remove the scale, lightly press on bottom of scale,

the top will lift up. Be extremely careful not to bend the

pointer (needle) or to move the Zero Adjust (ZA).

See figure II.G.7.

7. Grasp top and slide scale slowly out from top. Be

extremely careful not to bend the pointer (needle) or to

move the Zero Adjust (ZA). See figure II.G.7.

8. Reverse the above procedure using the new scale.

NOTE: While replacing the clear plastic cover, ensure the

ZA screw pin on the cover aligns with the inverted “Y” shaped

slot on the meter movement. See figure II.G.7.

IF, the inverted “Y” shaped slot is inadvertently moved -

BEFORE installing cover:

a. Re-center the inverted “Y” to the meter case as

shown in figure II.G.7.

b. Center the Zero Adjust (ZA) PIN on the clear

plastic cover as shown in figure II.G.8.

9. Using a standard slot screwdriver, ensure the ZA is

operating - pointer swings left and right when turning

the ZA with a standard slot screwdriver. See figure

II.G.9.

10. Set Meter ZERO as shown in figure II.G.9.

11. Reinstall meter to front panel, reversing steps 1-3

(except OEM models).

GASKET

Harness to Main

CB Connector

Backing Plate

METER

Mounting Screws (4)

BROWN WIRE

RED WIRE

+

CLEAR COVER

Pry Here Carefully

CASE

Figure II.G.5

Figure II.G.4

Figure II.G.6

!

CAUTION - READ FOLLOWING CAREFULLY