Myron L 729II User Manual

Page 38

Advertising
background image

34

D.

ALARM/CONTROL KIT

(Digital & OEM Models only)

-TH

Controller wiring Harness (requires second alarm/control

module) ordered with Monitor/controller.

THO

Controller wiring Harness (requires second alarm/control

module) ordered separately.

1. DESCRIPTION

The optional TH/THO Controller Kit adds more flexibility to

the Temperature Module by adding the ability to Alarm and/or

Control using the 24VDC solid state output to drive a PA - Piezo

electric alarm, and/or a 10 amp relay to operate a heater or

chiller as required.

Kit includes

THO kit comes with all items necessary to install and operate:

Alarm/control Harness (replaces second alarm/control relay har-

ness) with switch, bezel, cap and two O-rings (006 & 008); and

TEMPERATURE label (# LTEMP).

Requirements

-SC/SCO Alarm/Control Module (#2 set point/relay) option

-TP/TPO Temperature Module option

The TH/THO uses SC/SCO Second alarm/control Module on Main

circuit board [Conductivity/TDS, Resistivity, pH or mV (ORP)].

2. INSTALLATION

Briefly -

The SCH (switch and LED) harness is removed from the SC

and front panel, and discarded.

The THO harness is installed to the SC, TPO or other control-

lable option, and front panel.

A label is installed on the front panel.

WARNING: BEFORE STARTING, IF MONITOR/

CONTROLLER IS INSTALLED, ENSURE THE POWER

IS OFF. FAILURE TO DO SO COULD CAUSE DAMAGE

TO THE INSTRUMENT, AND COULD BE HARMFUL

OR FATAL TO PERSONNEL. ONLY QUALIFIED

PERSONNEL

SHOULD

INSTALL

OR

SERVICE

ELECTRICAL EQUIPMENT.

Physical

First, you must decide which location to install the THO harness

LED and Set Point switch. This is easily determined by how the

Primary or the Main CB Alarm/control is used. If the Primary

Alarm/control is used as a HIGH control, as in Conductivity, than

the THO harness is installed in the lower location - see section

III.D.2.a below.

If your Primary or Main circuit board Alarm/control is used as a

LOW control, as in Resistivity, than the THO harness is installed

in the HIGH location — see section III.D.2.b below.

WARNING: Before installing switch keep in mind there are two

(2) O-rings installed on each switch, one (1) on the threaded

shank and the other is under the push button. Both of these

must be reinstalled properly to maintain IP64/NEMA 3 ratings,

see figure III.D.4.

a. HIGH Primary Control

If the front panel has all ready been removed from the enclosure

skip to step 3.

1. Using a standard slot screwdriver remove the four (4)

screws on the front panel.

2. Carefully wiggle the front panel to loosen the and pull

gently toward you. Do not pull more than about 8

inches/20CM or you could damage the wiring harness.

3. Remove SCH harness from front panel and SC/SCO

Second Alarm/Control Module.

a. Carefully remove the RED LED from the front panel

!

CAUTION - READ FOLLOWING CAREFULLY

Figure III.D.1

FRONT PANEL

Rear View

REMOVE

PRIMARY

RED LED

REMOVE

PRIMARY SET

POINT SWITCH

Green

LED

LOW

REMOVE

SC/SCO

SET POINT

SWITCH

and

RED LED

FS

ZERO

DIS

SP2

HYS2

DISPLAY CONNECTION

DISPLAY

SELECT

TEMP

CONTROL

HARNESS

DISPLAY

BRN

BLU

+

0-5VD

-

INC

INC

DEC

DEC

TEMP

SENSOR

TP/TPO - TEMPERATURE MODULE

DEC DEC

BR

B

L

O

R

-

+

SPC

750

II

HIGH

LOW

SET POINT

MICROSIEMENS / CM

FRONT PANEL

Figure III.D.2

ROUND

BEZEL

NUTS

TEMPERATURE