H. 3 SENSOR / 3 RANGE MODULE

3SRO 3 Sensor / 3 Range Switch Module with 3 foot / 1 meter

cable ordered separately (customer installed and

mounted).

3SRE 3 Sensor / 3 Range Switch Module with 3 foot / 1 meter

cable AND separate Enclosure (as an accessory).

Enclosure is the same size as the Monitor/controller -

4.8”W x 6”H x 3.9”D.

1. DESCRIPTION

3 Sensor / 3 Range Switch Module is a circuit board (CB) with 2

rotary switches, and 3 groups of 5 sensor

connectors where up to three sensors may be connected.

The 3SR will function with all Conductivity/TDS and Resistivity

models, analog or digital. For ease of description, 3SR will be

used throughout, and in the following description each switch

function will be described separately.

The 3 Sensor Switch allows the user to view 3 separate sensor

locations and to set the alarm/control function using only one

Monitor/controller.

As shipped from the factory, the number “1“ Sensor position is

set to be the primary or the controlling sensor while the other

sensor positions are monitoring ONLY. This feature disables or

stops the alarm/control function from operating when changing

sensors. In other words, while in the #2 or #3 position the alarm/

controller is disabled NO ALARM or CONTROL. However, any

of the sensor positions many be selected to control by resetting

the select switches as shown on the CB and described below.

The 3 Range Switch allows the user to select which “range”

(Cond/TDS = X1, X10 or X100 / RES X1, X.1 or X.01) to view,

and to set for alarm/control function. The 3R may be configured

- via select switches - to disable or stop the alarm/control func-

tion from operating when changing ranges.

It is available in multipliers of “10” only.

Example #1 - Conductivity/TDS: 758II-117-3R = 500µS (X1)

w/ 5000µS (X10) & 50mS (X100) additional ranges.

Example #2 - Resistivity: 753II-11-3R = 20MΩ (X1) w/ 2MΩ

(X10/X.1) & 200KΩ (X100/X.01) additional ranges.

As shipped from the factory, the number “1” Range position is

the primary or the alarm/control position while the other posi-

tions are monitoring ONLY. In other words, while in the #2 (X.1)

or #3 (X.01) position the alarm/controller is disabled. In the

above example #2 the 20MΩ (X1) is the primary or alarm/control

range.

NOTE: It is NOT possible to have different alarm/control settings

for each sensor and range combination. The alarm/control func-

tion will always be the same as a percentage of full scale.

Specifications

NONE

2. INSTALLATION

Briefly -

The 3SR & 3SRO must be installed/mounted by the user out-

side of the Monitor/controller enclosure by drilling two (2) holes

in user panel. For the 3SRE, a cutout the size of the enclosure

is required.

A five conductor cable is connected to the 3SR Switch Module.

This cable is connected to the main Monitor/controller circuit

board (CB) at the sensor input connectors.

On a Conductivity Monitor/controller, the Range Module is

removed from Monitor/controller CB and re-installed on the 3SR

Switch Module as marked.

On a Resistivity Monitor/controller the Range Module is re-

moved from Monitor/controller CB and a special 3 Range-Range

Module is installed on the 3SR Switch Module as marked.

A cable harness from the 3SR Switch Module is connected to

the main Monitor/controller circuit board in place of the Range

Module.

A separate 2 wire connector & harness is for the alarm/control

disable or kill, and attaches to the small 2 pin male connector

S3 located to the left of the large input sensor connectors on

the main control CB. NOT INSTALLED ON MONITOR ONLY

MODELS.

WARNING: BEFORE STARTING, IF MONITOR/

CONTROLLER IS INSTALLED, ENSURE THE POWER

IS OFF. FAILURE TO DO SO COULD CAUSE DAMAGE

TO THE INSTRUMENT, AND COULD BE HARMFUL

OR FATAL TO PERSONNEL. ONLY QUALIFIED

PERSONNEL

SHOULD

INSTALL

OR

SERVICE

ELECTRICAL EQUIPMENT.

Physical

3SRO Switch Module only - A suitable location must be found

that is both convenient to the user plus allows sufficient clear-

ance for both switches and the CB to clear any obstructions.

3SRE - 3 Sensor / 3 Range Module with enclosure, a cutout the

size of the enclosure is required. Install enclosure per Operation

Manual and skip to #4 below.

1. Two 3/8” (9.5mm) holes must be drilled on 2” (26mm)

centers in the “user” panel. These holes must be

situated in a location that allows the 3SR Switch

Module to be mounted.

2. Install label, aligning on the drilled holes.

3. Mount the 3SR Switch Module to the panel with

hardware supplied.

4. Attach 5 wire cable harness from the 3SR Switch

Module to the main Monitor/controller CB at the sensor

input location as shown on connector label.

5. Remove the Range Module from the Main

Monitor/controller CB.

6. For Conductivity models, re-install Range Module onto

the 3SR Switch Module in the LEFT location as

marked.

For Resistivity models install replacement SPECIAL 3

Range-Range Module, part number RM-18 onto the

3SR Switch Module in the LEFT location.

7. Attach cable harness from the 3SR Switch Module

to the main Monitor/controller CB in the place of the

removed Range Module, as shown in figure III.H.1.

8. Attach the separate small cable harness from the 3SR

Switch Module to the small 2 pin male connector

located to the left of the sensor input connectors on

the main CB, as shown in figure III.H.1. NOT

INSTALLED ON MONITOR ONLY MODELS.

!

CAUTION - READ FOLLOWING CAREFULLY