Myron L 729II User Manual

Page 44

Advertising
background image

40

F.

3 SENSOR INPUT MODULE

3SO

3 Sensor Input Module with 3 foot / 1 meter cable

ordered separately (customer installed and

mounted).

3SE

3 Sensor Input Module with 3 foot / 1 meter cable

AND separate Enclosure (as an accessory).

Enclosure is the same size as the

Monitor/controller - 4.8”W x 6”H x 3.9”D.

1.

DESCRIPTION

3 Sensor Input Module is a circuit board (CB) with a 4 pole 3

position rotary switch, and 3 groups of 5 connectors where up

to three sensors may be connected. This option allows the user

to view 3 separate sensor locations and to set the alarm/control

function using only one Monitor/controller. The setpoints will not

change nor are they separately adjustable for each sensor.

As shipped from the factory, the number “1“ sensor position is

set to be the primary or the controlling sensor while the other

sensor positions are monitoring ONLY. This feature disables or

stops the alarm/control function from operating when changing

sensors. In other words, while in the #2 or #3 position the alarm/

controller is disabled. However, any of the sensor positions may

be selected to control by resetting the SELECT SWITCHES as

shown on the CB and described below. The 3S will function with

all Conductivity/TDS and Resistivity models, analog or digital.

Specifications

NONE

2. INSTALLATION

Briefly -

The 3SO must be installed/mounted by user outside of the

Monitor/controller housing by drilling a hole in user panel. For

the 3SE, a cutout the size of the enclosure is required.

A five conductor cable is connected to the 3S Module CB. This

cable is connected to the main Monitor/controller circuit board

(CB) at the sensor input connectors.

A separate small connector & harness is for the alarm/control

disable or kill, and attaches to the small 2 pin male connector

located to the left of the large sensor input connectors on the

main CB. NOT INSTALLED ON MONITOR ONLY MODELS.

WARNING: BEFORE

STARTING,

IF

MONITOR/

CONTROLLER IS INSTALLED, ENSURE THE POWER

IS OFF. FAILURE TO DO SO COULD CAUSE DAMAGE

TO THE INSTRUMENT, AND COULD BE HARMFUL

OR FATAL TO PERSONNEL. ONLY QUALIFIED

PERSONNEL

SHOULD

INSTALL

OR

SERVICE

ELECTRICAL EQUIPMENT.

Physical

3SO Module only - A suitable location must be found that is both

convenient to the user and allow sufficient

clearance for the 3 Sensor Switch Module to clear obstructions.

3SE - 3 Sensor Switch Module with enclosure, a cutout the size

of the enclosure is required. Install enclosure per Operation

Manual, and skip to #4 below.

1. A 3/8” (9.5mm) hole must be drilled in the “user” panel.

This hole must be situated in a location that allows the

3 Sensor Input Module to be mounted.

2. Install label, aligning on the drilled hole.

3. Mount the 3 Sensor Input Module to the panel with

hardware supplied.

4. Attach 5 wire cable harness from the 3 Sensor Input

!

CAUTION - READ FOLLOWING CAREFULLY

SENSOR

SWITCH

SENSOR 1

SENSOR 2

SENSOR 3

3SR MODULE

3S

ON

COND

RES

X1

X10

X100

S1

S2

S3

SENSOR 1

SENSOR 2

SENSOR 3

SELECT

SWITCH

TO SENSOR INPUT

SENSOR

OUTPUT

SENSOR INPUT

3 S

MAIN CIRCUIT BOARD

BACK/CONNECTOR SIDE

NOT INSTALLED ON

MONITOR ONLY MODELS

3 SENSOR INPUT MODULE (3S/3SO)

BK WT RD GN NU R- R+

BK WT RD GN NU

BK WT RD GN NU

BK WT RD GN NU

3 SR MODULE

SENSOR 1

SENSOR 2

SENSOR 3

SELECT SWITCH

ON

COND RES X1 X10 X100 S1 S2 S3

3

S

Figure III.F.1

SENSOR OUTPUT