4. Turn the front panel around so that the back side is

facing you and set aside for now.

5. Carefully remove front panel, leaving the harness

connected.

6. Using the enclosure cutouts, install the proper wire and

watertight cable restraint (not provided) to comply with

local electrical codes.

7. Neatly connect wires to the Main Circuit Board

connectors, as shown in figure II.E.1.

*CAUTION: The input power connectors require only a small

screwdriver or a pen to push on the release levers. The release

levers may be broken or damaged if not pushed straight toward

the circuit board. DO NOT push the release levers sideways.

2. 115/230 VAC CONVERSION

1. Before turning power on to the Monitor/controller

ensure the proper input voltage has been selected.

Failure to do so will blow the fuse. It could, under some

conditions, cause injury and damage the instrument

voiding the warranty.

2. Locate switch located next to the fuse holder.

3. Using a screwdriver, turn switch to required voltage.

3. CONNECTING THE SENSOR CABLE

For OEM models skip #1.

1. Place the Sensor’s interface cable and user supplied

watertight cable restraint into the enclosure’s

appropriate access hole.

2. Install the sensor cable wire to comply with local

electrical codes. Follow the color code as marked.

See figure II.E.1.

*CAUTION: The sensor connectors require only a small

screwdriver or a pen to push on the release levers. The release

levers may be broken or damaged if not pushed straight toward

the circuit board. DO NOT push the release levers sideways.

a. MODIFICATION FOR US PHARMACEUTICAL

25 (No Temperature Compensation)

This simple modification will allow your Monitor/controller

to meet the USP 25 requirements by defeating the normal

temperature compensation circuit thus giving “uncompensated”

readings as required.

Specifications:

As required to meet USP25.

Installation

Briefly -

For Conductivity/TDS, a resistor is installed in place of the

sensor “temperature” sensing leads.

For Resistivity, two resistors are installed in place of the sensor

“temperature” sensing leads.

The extra sensor leads are either cut off or the ends are

wrapped in tape to prevent shorting.

WARNING: BEFORE STARTING, IF MONITOR/

CONTROLLER IS INSTALLED, ENSURE THE POWER IS

OFF. FAILURE TO DO SO COULD CAUSE DAMAGE TO

THE INSTRUMENT, AND COULD BE HARMFUL OR FATAL

TO PERSONNEL. ONLY QUALIFIED PERSONNEL SHOULD

INSTALL OR SERVICE ELECTRICAL EQUIPMENT.

Requirements:

For Conductivity/TDS; one 10kΩ 1% resistor, user supplied or

may be ordered from the Myron L Company.

For Resistivity; one 100kΩ 1% resistor, and one 5.49kΩ

1% resistor, user supplied or may be ordered from the

Myron L Company.

NOTE: When opening instrument, remove front cover with care;

a ribbon cable connects the front panel and main board. If the

front panel has all ready been removed from the enclosure skip

to #4.

1. Using a standard slot screwdriver remove the four (4)

screws on the front panel.

2. Carefully wiggle the front panel to loosen and pull

gently toward you. Do not pull more than about 8

inches/20CM or you could damage the wiring harness.

3. Turn the front panel around so that the back side is

facing you and set aside.

4. For Conductivity/TDS Monitor/controllers;

a. If sensor is installed, locate and remove the RED

(RD) and the GREEN (GN) leads from MAIN

Circuit Board, as shown in figure II.E.2.

b. Cut off or tape RED (RD) and the GREEN (GN)

leads from sensor.

c. Install 10kΩ resistor at RED (RD) and the GREEN

(GN) connector locations, as shown in figure

II.E.3.

For Resistivity Monitor/controllers;

a. If sensor is installed, locate and remove the

GREEN (GN), RED (RD), and the NEUTRAL (NU)

leads from MAIN Circuit Board, as shown in figure

II.E.4.

b. Cut off or tape GREEN (GN), RED (RD), and the

NEUTRAL (NU) leads from sensor.

c. Install 100kΩ resistor at BLACK (BK) and GREEN

(GN) connector locations, as shown in figure

II.E.5.

12

Conductivity/TDS Main CB Assembly

Figure II.E.2

-121

2000µS

UP

}

HYS1

SP1

FS SW

BK WT RD GN NU R- R+

DIS

CAL

3S

FULL SCALE

PUSH TO TEST

INC

DEC

SPC

0-10VDC

OUTPUT

SENSOR LEADS

REMOVE THESE TWO LEADS

!

CAUTION - READ FOLLOWING CAREFULLY