For Conductivity/TDS - X1, X10 and X100.

For Resistivity - X1, 0.1, and 0.01.

If adding an alarm/control range.

1. Locate the 8 position select switch on the 3SR Module.

2. Set appropriate select switch UP or ON, i.e. to add

X100, set switch marked X100 UP or ON, see

figure III.H.7.

If changing alarm/control from one range to another, i.e. X1 to

X10 Range.

1. Locate the 8 position select switch on the 3SR Module.

2. Set the X1 select switch in the DOWN or OFF position.

3. Set the X10 select switch in the UP or ON position, see

figure III.H.8.

To disable either COND/TDS or Resistivity alarm/control

function completely.

1. Locate the 8 position select switch on the 3SR Module.

2. Set all X prefixed and S prefixed select switches in the

DOWN or OFF position, see figure III.H.9.

To reconfigure Conductivity to Resistivity or vise versa.

If you have a Resistivity Monitor/controller and ordered a 3SRO

Module as an accessory you may need to reconfigure the select

switch for it to Range properly.

1. Locate the 8 position select switch on the 3SR Module.

2. Set the COND select switch in the DOWN or

OFF position, see figure III.H.10.

3. Set the RES select switch in the UP or ON position,

see figure III.H.10.

Naturally, if you have a 3SR Module configured for a Resistivity

Monitor/controller and desire it to be for a Conductivity Monitor/

controller, reverse #2 & 3 above.

REASSEMBLY:

1. Carefully reinstall the front panel, bottom first. Ensure

no wires have been pinched between enclosure and

front panel.

2. Reinstall the four (4) screws and tighten.

3. To operate, turn power ON.

Figure III.H.6

Figure III.H.8

Figure III.H.7

Figure III.H.9

Factory Default

ON

ON

ON

COND RES X1 X10 X100 S1 S2 S3

COND RES X1 X10 X100 S1 S2 S3

COND RES X1 X10 X100 S1 S2 S3

COND RES X1 X10 X100 S1 S2 S3

ON

COND RES X1 X10 X100 S1 S2 S3

Figure III.H.10

ON

Advertising
This manual is related to the following products: