4. CONVERTING A CURRENT

TO A VOLTAGE

Current measuring devices actually measure voltage, but have

an internal resistor as shown in figures III.B.1 & III.B.2 and are

scaled to display in current. If you have a voltage input instru-

ment and you wish to utilize the current from the 4A (4-20mA)

Module, the following will help you to make this conversion.

To convert a 4-20 mA current output to operate or drive a volt-

age input device it is necessary to install a LOAD resistor across

the input terminals as shown in figures III.B.6 & III.B.7. The

value of the resistor is chosen to match the input voltage range,

i.e. 0-10 Volts requires a 500Ω resistor and will produce a 2 to

10 input voltage. This floating zero is useful to indicate a broken

4-20 input wire when the indication is zero volts.

For other input ranges, divide the input voltage range by 0.02,

the answer will be in ohms.

NOTE: The tolerance of the load resistor directly affects the

accuracy of the resulting voltage, i.e. 5% resistor = 5% error.

26

VOLTAGE INPUT INSTRUMENT

+ 35 VDC

MAXIMUM

VOLTAGE INPUT INSTRUMENT

Figure III.B.7

(+)

(-)

+

-

(+)

SIGNAL OUT

POWER OUT

POWER IN

+

Remote - Powered

EXTERNAL LOAD

RESISTOR

Figure III.B.6

(+)

(-)

+

SO PO PI

SO PO PI

-

(+)

SIGNAL OUT

POWER OUT

POWER IN

+

Self - Powered

N

EXTERNAL LOAD

RESISTOR

N

For an input voltage range of:

10.0 Volts

the resistor value is 500 ohms

indicates

2.0 V @ 4 mA

5.0 Volts

250 ohms

1.0 V @ 4 mA

1.0 Volts

50 ohms

0.2 V @ 4 mA

100 millivolts

5 ohms

20 mV @ 4 mA

10 millivolts

0.5 ohms

2 mV @ 4 mA

Advertising