Myron L 729II User Manual

Page 4

Advertising
background image

connected to an appropriate Sensor, and that the Sensor is

immersed in water within the range that the Monitor/controller will

be required to read; and the front panel is removed.

1. Make a note of the reading on the display.

2. While pressing the Calibration/Full Scale Test Switch

(FSSW), verify that the front panel display is indicating

a full scale reading. If not, see Calibration, section V.C.

3. Press and hold the “SET POINT” switch on the front

panel. Using a tweaker or a small screwdriver, adjust

the Set Point trimmer adjustment screw on the circuit

board to sweep the display from zero to full scale. (A

digital display may be blank at the full scale end. This

is normal.) Listen for the alarm relay to click on and off

as the alarm set point moves past the water reading.

4. Adjust the alarm to the desired set point value. Release

the “SET POINT” switch.

NOTE: For Models with SC/SCO module, repeat STEPS 3 & 4 to

check out Set Point #2.

QUICK SET POINT ADJUSTMENT -

See Section IV.C.3.

The set point setting is based upon the user’s particular water

purity specifications or requirements. NOTE: The optional

second relay/alarm is “stacked” on the first relay/alarm,

therefore, when setting the optional second relay/alarm Set

Point, the #1 Set Point must be ‘set’ first.

1. While pressing the “SET POINT” switch, turn the Set

Point #1 adjustment screw (see figure V.A.1) until the

desired set point value is indicated on the display.

HYSTERESIS (DEAD BAND) ADJUSTMENT -

See Section IV.C.4.

PRIMARY COMPONENT IDENTIFICATION -

Section V.A.

Review the figure below to familiarize yourself with the Main

circuit board assembly. The diagram has the second alarm/

control module option installed.

QUICK CALIBRATION - Section V.C.

WARNING: When performing calibration procedures,

the technician must take extreme care to avoid

contacting the circuitry other than the CALibration

control. Failure to do so could result in damage to

the equipment, property and/or personal injury.

The following assumes the front panel has been

removed and the power is ON.

ELECTRONIC CALIBRATION (CIRCUIT ONLY) -

See Section V.C.1.

Full Scale Adjustment

1. Press and hold the Full Scale Test switch. The display

should indicate Full Scale for the particular range

selected, i.e. 0-500 ppm should indicate 500. If not, set

to Full Scale with the CALibration control.

2. Turn power OFF.

3. Re-install front panel as described in “REASSEMBLY”.

4. To operate, turn power ON.

10VDC Calibration - See Section V.C.1.b.

Using Standard Solutions - Section V.C.2.

The BEST method of verifying and recalibrating your

conductivity/TDS Monitor/controller is with NIST traceable

Standard Solution (available from the Myron L Company).

Because it includes the sensor, the entire system is recalibrated.

NOTE: Since standard solution calibrations are NOT practicable

with resistivity models, another means of verification or

calibration of resistivity models is to use the transfer standard

method, using a hand-held or portable instrument capable of

resistivity measurements, i.e. the Myron L Ultrameter II™. See

section V.C.4 for description.

The following procedure describes the easiest method for

standard solution calibration of your Conductivity/TDS

Monitor/controller.

1. Using a standard solution which is 60-90% of full scale

of the instrument, rinse thoroughly and fill a clean glass

beaker with the standard solution.

2. Place sensor in the beaker of standard solution. The

level of standard solution must be high enough to cover

at least 1/2” above cross hole.

3. Slowly shake the sensor to remove air bubbles from

inside the sensor bore hole.

4. Allow 5-10 minutes for temperature to equilibrate. For

the quickest and the best results, both the sensor and

solution should be at the same temperature.

5. Turn power ON.

6. If the reading is different from the standard solution,

adjust CALibration control on the main circuit board

until the reading matches the solution value.

7. After adjustment, turn power OFF.

8. Re-install front panel as described in “REASSEMBLY”.

9. To operate, turn power ON.

SENSOR SUBSTITUTE CALIBRATION -

See Section V.C.3.

TRANSFER STANDARD METHOD - See Section V.C.4.

B

L

K

W

H

T

SENSOR

RANGE

MODULE

MAIN

CALIBRATION

CONTROL

SC/SCO

OPTIONAL

SECOND

ALARM/

CONTROL

MODULE

#2 RELAY

LIGHTS &

SWITCH

CONNECTOR

RUBBER TAPE

(DO NOT

REMOVE)

RANGE

MODULE

ALIGNMENT

3SO/3SE

&/OR

3RO/3RE

SWITCH

OR

FLOW SWITCH

CONNECTION

POWER

AC LINE/ +DC

AC NEUTRAL/ -DC

GROUND

RELAY #1

}

COM

NC

NO

DISPLAY/METER

CONNECTOR

4-20 CONNECTOR

SET POINT #1

CONVERSION

SET POINT #2 CONTROL

SET POINT #2

CONVERSION

DISPLAY CALIBRATION

CONTROL (FACTORY SET)

SET POINT #1

HYSTERESIS

RIGHT INC

LEFT DEC

CHASSIS GROUND - OEM

SET POINT #2 HYSTERESIS

RIGHT INC / LEFT DEC

SOLID STATE (24VDC 30mW)

OUTPUT

PA™ PIEZO ELECTRIC ALARM

OR

RA™ REMOTE ALARM OR

CUSTOMER CONNECTION

SOLID STATE (24VDC 30mW)

OUTPUT

RA™ REMOTE ALARM AND

PA™ PIEZO ELECTRIC ALARM

OR

CUSTOMER CONNECTION

RELAY #2

}

COM

NO

NC

FULL SCALE

PUSH TO TEST

SET POINT

#1 CONTROL

0-10VDC OUTPUT

N

E

U

(+)

(-)

R

E

D

G

R

N

3S

PA

HYS1

SP1

FS SW

DIS

INC

DEC

CAL

INC

-121

2000µS

}

SPC

TRANSFORMER

FUSE*

115/

230

751 756

752 757

753 758

754 759

CHS

GND

PWR

C GD NC NO CM

CM

N

O

N

C

SPC

INC

UP

+

+

PA

RA

{

BK WT RD GN NU R- R+

REMOVE T

O INST

ALL

SECOND RELA

Y

5/21/03

Figure V.A.1

Advertising
This manual is related to the following products: