alarm for approximately three minutes or until you

improve the water quality (readjust controller set point).

If three minutes muting is fine for your application, skip

to #10.

10. If three minutes is too long or too short, adjust control

inside RA until desired mute time is achieved (adjustable

from approximately 15 seconds to 10 minutes). See

figure III.J.1.

11. Replace the bottom of the RA, and secure to surface

you have selected for its installation.

12. Carefully reinstall the controller front panel, bottom first.

Ensure no wires have been pinched between enclosure

and front panel.

13. Reinstall the four (4) screws and tighten.

14. To operate, turn power ON.

*NOTE: If the RA does not sound off;

1. Check the polarity of the extension wire connections.

2. Be sure the controller is actually switching (relay will

click).