Click the calibrate button, Select “using the barco lcd sensor” and click ok – Barco E1 User Manual

Page 77

Advertising
background image

NioWatch operation

78

System manual E1

option in the Application Settings dialog is set. See
“Calibration tab” in the description of the Application
Settings dialog.

2. If you have multiple displays connected (multi-head system),

select the display you wish to calibrate by clicking on the
corresponding numbered icon in the Display Settings dialog.

3. Click the Calibrate button.

4. A message appears, showing the different calibration

options.

Figure 39: Calibration choices

5. Select “Using the BARCO LCD sensor” and click OK.

This option is grayed in case NioWatch does not find a
connected Barco LCD sensor.

6. Follow the guidelines on the screen to complete the

calibration successfully.

Note:

If the Equalization option is set in the Application Settings
and NioWatch finds at least one other display of the same
type that is calibrated, the “Match with reference” page
appears during the calibration process.