When done, click button set point 1 of 20, Click ok to finish, To reset to the latest calibration – Barco E1 User Manual

Page 79

Advertising
background image

NioWatch operation

80

System manual E1

Figure 41: DICOM Optimizer dialog

3. Adjust the slider until there is almost no visible difference

between the background of the image above the slider and
the bitmap inside this image.

4. When done, click button Set point 1 of 20.

5. Repeat this procedure until the button is grayed. This

indicates the last point is set.

6. Click OK to finish.

To reset to the latest calibration:

1. Follow step 2 and 3 of the calibration with Barco LCD sensor

(see page 77).

2. In the Calibration dialog (see step no. 4. in the description of

the calibration with sensor), check the option Use the latest
calibration
.