Operating manual – Eppendorf 5810 R Centrifuge User Manual

Page 14

Advertising
background image

Centrifuge 5804/5804 R/5810/5810 R — Operating manual

14

Blood collection tube

5810 748.003

flat

3,000 × g

3 to 15 mL

Ø 16 mm

4,000 rpm

16/64

108 mm/108 mm

16.8 cm

Vessel

5810 721.008

flat

3,000 × g

7 to 17 mL

Ø 17.5 mm

4,000 rpm

16/64

118 mm/118 mm

16.8 cm

Conical tube

5810 722.004

conical

3,100 × g

15 mL

Ø 17.5 mm

4,000 rpm

12/48

119 mm/121 mm

17.3 cm

Conical tube

5810 723.000

conical

3,100 × g

50 mL

Ø 31 mm

4,000 rpm

5/20

116 mm/122 mm

17.3 cm

Centriprep

5810 739.004

flat

3,100 × g

50 mL

Ø 31 mm

4,000 rpm

5/20

(Do not use the
aerosol-tight cap.)/
121 mm

17.3 cm

Conical tube, skirted

5810 739.004

5804 737.008

flat

3,100 × g

50 mL

Ø 31 mm

4,000 rpm

5/20

(Do not use the
aerosol-tight cap.)/
119 mm

17.3 cm

Bottles

5825 722.000

flat

3,100 × g

180 to 250 mL

Ø 62 mm

4,000 rpm

1/4

(Do not use the
aerosol-tight cap.)/
133 mm

17.3 cm

Wide-neck bottle

5810 728.002

flat

3,220 × g

400 mL

Ø 81 mm

4,000 rpm

1/4

(Do not use the
aerosol-tight cap.)/
133 mm

18.0 cm

Vessel

Vessel

Adapters

Adapter bottom
shape

Max. g-force

Capacity

Order no.

(international)

Tube diameter

Max. speed

Vessels per
adapter/rotor

Max. tube length
with/without
aerosol-tight
bucket cap

Centrifugation radius

Eppendor

f

EN

Operating manual

Advertising
This manual is related to the following products: