3 installing the instrument, Operating manual – Eppendorf 5810 R Centrifuge User Manual

Page 49

Advertising
background image

49

Centrifuge 5804/5804 R/5810/5810 R — Operating manual

4.3

Installing the instrument

Perform the following steps in the sequence described.

1. Allow the device to warm up for at least 3 hours (5804/5810) or 4 hours (5804 R/5810 R) to

the ambient temperature to prevent damage to the electronic components from condensation
and damage to the compressor (only5804 R/5810 R).

2. Check that the mains voltage and frequency match the requirements on the device type plate.

Centrifuge 5804 R/5810 R with power supply voltage 120 V: See also the notes on the power
supply at the end of this chapter.

3. Connect the centrifuge to the mains and switch it on using the mains power switch on the

right side of the device.

The key open lights up.

Display is illuminated.

4. Open the centrifuge lid using the open key.
5. Use the details included in the scope of delivery to check that the delivery is complete

(see Delivery package on p. 11).

6. Check all parts for any transport damage. Contact your dealer if any damage is found.
7. Only 5804 R/5810 R: Insert the condensation water tray at the front of the device into the

provided holder (

1

3).

Tab.

1:

Centrifuge 5804 R / 5810 R with mains voltage 120 V in two versions

WARNING!

Risk from incorrect supply voltage

Only connect the device to voltage sources that match the electrical requirements listed on
the name plate.

Only use sockets with a protective earth (PE) conductor and suitable power cable.

NOTICE!

Centrifuge 5804 R/5810 R: compressor damage after improper transport.

Only switch on the centrifuge 4 hours after installation.

15 A IEC power cable

20 A variant

Conventional IEC power cable.

Connection to standard socket (120 V/
15 A).

Standard cooling performance:

– Increased minimum achievable

temperatures at maximum speed of
centrifugation.

– Slower cooling down to set temperature.

Power cable fitted permanently to the
device.

Special mains connection required (120 V/
20 A).

Increased cooling performance.

– Lower temperatures at maximum speed

of centrifugation possible.

– Quicker cooling down to set

temperature.