Centrifuge 5804/5804 R/5810/5810 R — Operating manual

2.4.11 Rotor S-4-104 (only 5810/5810 R)

Max. g-force:

3,214 × g

Rotor S-4-104

Round bucket
750 mL

Aerosol-tight cap

Max. speed:

3,900 rpm

Swing-bucket rotor
with 4 × 750 mL
round buckets

Max. load per bucket
(adapter, tube and
contents):

1,000 g

Vessel

Vessel

Adapters

Adapter bottom
shape

Max. g-force

Capacity

Order no.

(international)

Tube diameter

Max. speed

Vessels per
adapter/rotor

Max. tube length
with/without
aerosol-tight
bucket cap

Centrifugation radius

Vessel

5825 740.009

continuous

3,197 × g

1.5/2 mL

Ø 11 mm

3,900 rpm

62/248

39 mm

18.8 cm

Vessel

5825 739.000

continuous

3,214 × g

5 mL

Ø 17 mm 3,900 rpm

14/56

60 mm

18.9 cm

Vessel

5825 738.004

round

3,044 × g

4 to 8 mL

Ø 13 mm × 100 mm

3,900 rpm

23/92

108 mm/115 mm

17.9 cm

Vessel

5825 736.001

round

3,061 × g

7,5 to 12 mL

Ø 16 mm × 98 mm

3,900 rpm

20/80

114 mm/119 mm

18 cm

Vessel

5825 736.001

round

3,061 × g

8 to 16 mL

Ø 16 mm

3,900 rpm

7/28
Load inner bores
only (Fig. 5 on p. 54).

(Do not use the
aerosol-tight cap.)/
125 mm

18 cm

Vessel

5825 743.008

round

3,044 × g

9 mL

Ø 17.5 mm ×
100 mm

3,900 rpm

20/80

106 mm/111 mm

17.9 cm

Conical tube

5825 734.009

conical

3,197 × g

15 mL

Ø 17 mm × 104 mm

3,900 rpm

14/56

120 mm/125 mm

18.8 cm

EN