4 information on product liability, Operating manual – Eppendorf 5810 R Centrifuge User Manual

Page 42

Advertising
background image

Centrifuge 5804/5804 R/5810/5810 R — Operating manual

42

For all other rotors and rotor lids of this centrifuge there is no service life limit as long as the
following requirements are met:

Correct use,

Recommended maintenance

Undamaged condition

The date of manufacture is stamped on the rotors in the format 03/12 (= March 2012) or on the
inside of the plastic rotor lids in the form of a clock .
To ensure aerosol tightness, the following applies:

Replace aerosol-tight rotor lids and caps after 50 autoclaving cycles.

Replace the seal of QuickLock rotor lids after 50 autoclaving cycles.

3.4

Information on product liability

In the following cases, the designated protection of the device may be compromised. Liability for
any resulting property damage or personal injury is then transferred to the operator:

The device is not used in accordance with the operating manual.

The device is used outside of its intended use.

The device is used with accessories or consumables which are not recommended by
Eppendorf.

The device is maintained or repaired by people not authorized by Eppendorf.

The user makes unauthorized changes to the device.

Rotor

Maximum service from commissioning onward

A-2-DWP-AT

100,000 mechanical cycles

7 years

A-2-DWP

34,000 mechanical cycles

7 years

A-4-44

34,000 mechanical cycles

7 years

A-4-62

40,000 mechanical cycles

7 years

A-4-81

100,000 mechanical cycles

7 years

F-34-6-38

75,000 mechanical cycles

7 years

FA-45-6-30

7 years

FA-45-48-11

75,000 mechanical cycles

7 years

FA-45-20-17

75,000 mechanical cycles

7 years

F-35-48-17

75,000 mechanical cycles

7 years

S-4-72

60,000 mechanical cycles

7 years

S-4-104

100,000 mechanical cycles

7 years

T-60-11

7 years

Accessories

Maximum service from commissioning onward

Aerosol-tight rotor lid, without replaceable
seals

50 autoclaving cycles

Rotor lid QuickLock

3 years

Seals of the QuickLock rotor lid

50 autoclaving cycles

Rotor lid and caps made of polycarbonate
(PC), polypropylene (PP) or polyetherimide
(PEI)

50 autoclaving cycles

3 years

Adapters

1 year