5 rotor a-2-dwp, Operating manual – Eppendorf 5810 R Centrifuge User Manual

Page 26

Advertising
background image

Centrifuge 5804/5804 R/5810/5810 R — Operating manual

26

2.4.5

Rotor A-2-DWP

Hint!

Check the load if using two fully loaded DWP plates.

Max. g-force:

2,250 × g

Rotor A-2-DWP

Deepwell plate
carrier

Max. speed:

3,700 rpm

Swing-bucket rotor with 2 Deepwell plate
carriers

Max. load per bucket
(adapter, plate and
contents):

380 g

Plate

Plate

Adapters

Adapter bottom
shape

Max. g-force

Capacity

Order no.

(international)

Max. speed

Plates/slides per
adapter/rotor

Max. loading height Centrifugation radius

Micro test plate

SBS adapter*

5825 718.003

flat

2,250 × g

96/384 wells

4,000 rpm

4/8

89 mm

14.7 cm

Cell culture plate

SBS adapter*

5825 718.003

flat

2,250 × g

4,000 rpm

4/8

89 mm

14.7 cm

Deepwell plate

SBS adapter*

5825 718.003

flat

2,250 × g

96 wells

4,000 rpm

2/4

89 mm

14.7 cm

Kit

SBS adapter*

5825 718.003

flat

2,250 × g

4,000 rpm

1/2

89 mm

14.7 cm

Tube in IsoRack

5825 708.008

flat

2,050 × g

24 × 0.5 mL

Ø 6 mm

3,700 rpm

1/2

89 mm

13.8 cm

Tube in IsoRack

5825 709.004

flat

1,990 × g

24 × 1.5/2 mL

Ø 11 mm

3,700 rpm

1/2

89 mm

13.3 cm

EN