7 rotor f-34-6-38, Operating manual – Eppendorf 5810 R Centrifuge User Manual

Page 29

Advertising
background image

29

Centrifuge 5804/5804 R/5810/5810 R — Operating manual

2.4.7

Rotor F-34-6-38

Vessel

5820 728.000

Round

16,233 × g

5.5 mL – 10 mL

Ø 16 mm

19,246 × g

1/6

-

12.0 cm

Vessel

5820 727.004

Round

16,233 × g

7.5 to 12 mL

Ø 16.4 mm

19,246 × g

1/6

119 mm

12.0 cm

Vessel

5820 729.007

Round

16,233 × g

9 mL

Ø 16.4 mm

19,246 × g

1/6

112 mm

12.0 cm

Rotor F-34-6-38
Fixed-angle rotor for
6 x 85 mL tubes

Max. g-force:

15,557 × g (5810 R: 18,514 × g)

Max. speed:

11,000 rpm (5810 R: 12,000 rpm)

Max. load (adapter,
tube and contents):

6 × 125 g

Vessel

Vessel

Adapters

Adapter bottom
shape

Max. g-force at 11,000 rpm
(5804/5804 R/5810)

Capacity

Order no.

(international)

Tube diameter

Max. g-force at 12,100 rpm
(5810 R)

Vessels per
adapter/rotor

Max. tube length
with rotor lid

Centrifugation radius

Vessel

5804 770.005

Round

15,300 × g

1.5/2 mL

Ø 11 mm

18,200 × g

4/24

43 mm

11.3 cm

Vessel

5804 777.000

conical

14,150 × g

16,842 × g

5 mL

Ø 17 mm

10,45 cm

1/6

-

Blood collection tube

5804 738.004

Round

14,339 × g

2 mL to 5 mL Ø

13 mm

17,065 × g

3/18

80 mm

10.6 cm

Vessel

Vessel

Adapters

Adapter bottom
shape

Max. g-force at 11,000 rpm
(5804/5804 R/5810)

Capacity

Order no.

(international)

Tube diameter

Max. g-force at 12,100 rpm
(5810 R)

Vessels per
adapter/rotor

Max. tube length
with rotor lid

Centrifugation radius

EN

Operating manual

Advertising
This manual is related to the following products: