Operating manual – Eppendorf 5810 R Centrifuge User Manual

Page 73

Advertising
background image

73

Centrifuge 5804/5804 R/5810/5810 R — Operating manual

Deceleration times according to DIN 58 970

Tab.

2:

Approximate deceleration times of the different rotors for the levels 0 to 9 (in seconds)
for 230 V devices

These values are to be considered as guidelines. Level 9 means "strongest braking", level 0
means "free deceleration". Considerable fluctuations can occur depending upon the condition of
the device and the load. The deceleration times for the 230 and 120 V devices are almost
identical.

5804/

5804 R

5810/5810

R

Rotor

0

1

2

3

4

5

6

7

8

9

-

•

A-4-81

532

189

174

143

131

109

95

85

59

31

-

•

A-4-81- MTP/

Flex

643

191

174

142

131

110

94

83

58

30

-

•

A-4-62

740

190

170

140

130

110

95

85

55

26

-

•

A-4-62-MTP

620

190

170

140

130

110

95

85

55

26

•

•

A-4-44

470

300

270

220

200

140

100

75

45

23

-

•

A-2-DWP-AT

857

231

202

176

157

135

μL

112

102

69

39

•

•

A-2-DWP

304

174

130

118

100

75

51

44

32

14

•

•

FA-45-6-30

759

423

322

231

205

178

148

113

93

58

•

•

F-34-6-38

880

370

280

190

170

150

125

95

75

54

•

•

FA-45-30-11

240

140

70

45

35

30

25

22

19

18

•

•

F-45-30-11

240

140

70

45

35

30

25

22

19

18

•

•

F-45-48-PCR

169

119

60

41

31

26

22

19

17

16

•

•

T-60-11

800

280

140

95

70

55

45

40

36

36

-

•

S-4-104

680

192

169

147

130

112

68

46

37

32

•

•

S-4-72

360

238

191

168

148

117

56

32

21

15

•

•

FA-45-48-11

454

269

141

98

76

62

46

40

35

31

•

•

FA-45-20-17

550

342

178

123

97

80

58

51

45

40

•

•

F-35-48-17

16

26

40

72

140

185

211

243

304

228

EN

Operating manual