8 troubleshooting, 1 resetting the excess current switch, 2 general errors – Eppendorf 5810 R Centrifuge User Manual

Page 67: 3 error messages, Operating manual

Advertising
background image

67

Centrifuge 5804/5804 R/5810/5810 R — Operating manual

8

Troubleshooting

8

Troubleshooting

If you cannot remedy an error with the recommended measures, please contact your local
Eppendorf partner. The contact address can be found online at: www.eppendorf.com/worldwide.

8.1

Resetting the excess current switch

The 230 V and 120 V devices have built-in thermal excess-current switches which function as
(all-pole) fuses. When the overload protection is actuated, these switch the power switch to OFF,
but do not switch it on again automatically.
To switch on the excess current switch again, proceed as follows:

1. Switch off the centrifuge using the power switch.
2. Wait for at least 20 seconds and switch on the centrifuge again.
The excess current switch will be automatically reactivated and the centrifuge is ready for
operation.

8.2

General errors

8.3

Error messages

If one of the following error messages appears, proceed as follows:
1. Remove fault (see Remedy).
2. Press open key to clear the error message.
3. If necessary, repeat centrifugation.
Some errors can have various causes. The actual cause is described in the message in the
device display.

Symptom/message

Cause

Remedy

No display.

No mains power connection.

Check the mains power connection.

No display.

Power failure.

Check the mains fuse of the device
(see Resetting the excess current switch
on p. 67)
.

Check the mains fuse of the laboratory.

Lid of the device cannot
be opened.

Rotor is still running.

Wait for rotor to stop.

Lid of the device cannot
be opened.

Power failure.

1. Check the mains fuse of the device

(see Resetting the excess current switch
on p. 67)
.

2. Check the mains fuse of the laboratory.
3. Activate the emergency lid release

(see p. 69).

Clean rotor

200 runs.

Clean the rotor and chamber (see p. 63).

Centrifuge brakes during
a short run centrifugation,
although the short key is
pressed.

The short key was released briefly more than
twice (protective function for the drive).

Press the short key continuously during a
short run centrifugation.

Symptom/message

Cause

Remedy

No rotor

Centrifuge does not start
up.

No rotor.

Insert the rotor.

EN

Operating manual