Operating manual – Eppendorf 5810 R Centrifuge User Manual

Page 23

Advertising
background image

23

Centrifuge 5804/5804 R/5810/5810 R — Operating manual

Conical tube

5804 717.007

conical

4,400 × g

15 mL

Ø 17.5 mm

5,000 rpm

2/8

121 mm/121 mm

15.7 cm

Vessel

5804 756.002

flat

4,200 × g

7 to 18 mL

Ø 20 mm

5,000 rpm

4/16

104 mm/107 mm

15.0 cm

Vessel

5804 757.009

flat

4,200 × g

18 to 30 mL

Ø 26 mm

5,000 rpm

2/8

100 mm/110 mm

15.0 cm

Conical tube

5804 758.005

conical

4,300 × g

50 mL

Ø 31 mm

5,000 rpm

1/4

(Do not use the
aerosol-tight cap.)/
122 mm

15.5 cm

Conical tube

5804 718.003

conical

4,400 × g

50 mL

Ø 31 mm

5,000 rpm

1/4

119 mm/122 mm

15.7 cm

Conical tube

5804 706.005

Max. load 144 g

(insert, tubes and

contents)

flat with conical
insert

4,500 × g

50 mL

-

5,000 rpm

-/8

(Do not use the
aerosol-tight cap.)/
120 mm

16.1 cm

Vessel

5804 759.001

flat

4,200 × g

30 to 50 mL

Ø 31 mm

5,000 rpm

1/4

108 mm/122 mm

15.0 cm

Conical tube, skirted

5804 759.001

5804 737.008

flat

4,200 × g

50 mL

Ø 31 mm

5,000 rpm

1/4

108 mm/122 mm

15.0 cm

Vessel

Vessel

Adapters

Adapter bottom
shape

Max. g-force

Capacity

Order no.

(international)

Tube diameter

Max. speed

Vessels per
adapter/rotor

Max. tube length
with/without
aerosol-tight
bucket cap

Centrifugation radius

EN