Operating manual – Eppendorf 5810 R Centrifuge User Manual

Page 54

Advertising
background image

Centrifuge 5804/5804 R/5810/5810 R — Operating manual

54

Abb. 5:

Loading the adapter 16 mm 75 - 100 mm with tubes >119 mm in length.

Fig.

5:

Loading the adapter 16 mm 75 - 100 mm with tubes >119 mm in length.

Abb. 6:

Symmetrical loading of the plates.

Fig.

6:

Symmetrical loading of the plates.

The plate arrangement shown on the right-hand side is incorrect, as the buckets will not
swing properly.

The same principle also applies to the loading of rotor A-4-81-MTP/Flex with 4 plates.

The plates have some play in the buckets.

Abb. 7:

Mixed loading of rotors

Fig.

7:

Mixed loading of rotors

8. Check the loading of the bucket.

Rotor

Mixing equipment

S-4-104

2 plate buckets (open bucket or plate
bucket)

2 round buckets

A-4-81/A-4-81-MTP/Flex

2 plate buckets (MTP or DWP bucket)

2 buckets for conical tubes

2 rectangular buckets

A-4-44

2 rectangular buckets

2 buckets for conical tubes

NOTICE!

Rotor damage due to mixed loading.
If you load the rotors A-4-62 and A-4-62-MTP with a mixed equipment, the rotors are damaged
during centrifugation.

Load all positions of the rotors A-4-62 and A-4-62-MTP with the same buckets.

Always load all 4 positions of the swing-bucket rotors.

EN

Operating manual