Centrifuge 5804/5804 R/5810/5810 R — Operating manual

44

3.5.2

Incorrect handling of the centrifuge

WARNING!

A defective gas spring is an insufficient support for the centrifuge lid.
A defective gas spring is an insufficient support for the centrifuge lid.

Make sure that the centrifuge lid can be opened completely and that it will remain in this
position.

Regularly check the gas spring for its proper function.

Have defective gas springs replaced immediately.

We recommend that the gas springs be replaced every 2 years by a service technician.

CAUTION!

Poor safety due to incorrect accessories and spare parts.
The use of accessories and spare parts other than those recommended by Eppendorf may
impair the safety, functioning and precision of the device. Eppendorf cannot be held liable or
accept any liability for damage resulting from the use of incorrect or non-recommended
accessories and spare parts, or from the improper use of such equipment.

Only use accessories and original spare parts recommended by Eppendorf.

NOTICE!

Damage to device due to spilled liquids.

1. Switch the device off.
2. Disconnect the device from the power supply.
3. Carefully clean the device and the accessories in accordance with the cleaning and

disinfection instructions in the operating manual.

4. If a different cleaning and disinfecting method is to be used, contact Eppendorf AG to ensure

that the intended method will not damage the device.

NOTICE!

Damage to electronic components due to condensation.
Condensation can form inside the device after the device has been moved from a cool to a
warmer environment.

5804/5810: Wait for at least 3 hours before connecting the device to the power supply.

5804/5810 only: : Alternative: Let the device heat up for 30 minutes right before a brief
transport.

5804 R/5810 R: Wait for at least 4 hours before connecting the device to the power supply.

NOTICE!

Centrifuge 5804 R/5810 R: compressor damage after improper transport.

Only switch on the centrifuge 4 hours after installation.

NOTICE!

Buckets swinging out in the wrong direction.
When working with the rotor S-4-104, using the wrong adapters with 500 mL Corning tubes may
lead to the loss of samples or damage to the centrifuge.

Only use the Eppendorf adapter intended for this purpose.

NOTICE!

Damage from knocking against or moving the device during operation.
If the rotor bangs against the rotor chamber wall, it will cause considerable damage to the device
and rotor.

Do not move or knock against the device during operation.