2 modbus holding registers, Modbus holding registers -5, Table 5 – CiDRA Passive Sonar Transmitters User Manual

Page 20

Advertising
background image

3.2

Modbus Holding Registers

Table 5

Modbus Holding Registers

Address Size

Type Value

Description

Values

1

1

Char

Password Input

Password entry, when passwords are configured.

0 – 65535.

4 1

Char

Write

Control

Controls ability to write and commit changes to
transmitter configuration.

Write 0x55AA to enable write
access, write 0xEDF1 to commit
changes.

10

1

Char

Set Password 0

Sets Password 0.

0 – 65535.

11

1

Char

Set Password 1

Sets Password 1.

0 – 65535.

12

1

Char

Set Password 2

Sets Password 2.

0 – 65535.

20 1

Char

Reset

Totalizers

Resets

all totalizers to zero.

Any Write.

21

1

Char

Clear Alarm

Clear any existing alarms.

Any Write.

22

1

Char

Reset Data History

Clears the data history memory.

Any Write.

1001 1

Char

PIPE_DIAM_SELEC
T

Selects method used to set pipe dimensions.

0 = ID/Wall (Uses
DISP_PIPE_DIAM and
WALL_THICKNESS), 2 =
Size/Sched (Uses PIPE_SS_SIZE
and PIPE_SS_SCHED), 3 =
OD/Wall (Uses PIPE_OD and
WALL_THICKNESS).

1002 1

Char

PIPE_DIAM_UNITS

Selects

units

used for 'Pipe ID'.

0 = Inches, 1 = millimeters

1003 1

Char

PIPE_OD_UNITS

Selects

units

used

for 'Pipe OD'.

0 = Inches, 1 = millimeters

1004 1

Char

PIPE_SS_SIZE

Selects pipe size. Will only be applied if 'Size /
Schedule' is selected for 'Pipe Diameter Input Mode'.

0=2,1=2.5,3,3.5,5,6,8,10,12,14,16,
18,20,22,24,26,28,30,32,34,36

1005 1

Char

PIPE_SS_SCHED

Selects pipe schedule. Will only be applied if 'Size /
Schedule' is selected for 'Pipe Diameter Input Mode'.

0=5S,1=10,10S,20,30,40,40S,60,8
0,80S,STD,XS

1006 1

Char

SOS_PIPE_WALL_T
HICKUNITS

Selects units used for 'SOS Pipe Wall Thickness'.

0 = Inches, 1 = millimeters

1007 1

Char

SOS_PIPE_MODUL
US_SEL

Selects either a pre-defined modulus or the option to
enter a custom value. Select 'Custom' to enter a value in
'SOS Pipe Modulus'.

0 = 1.9305e8 kPa (SS), 1 =
2.0684e8 kPa (Steel), 2 =
3.4473e6 kPa (PVC), 3 = Custom

1008 1

Char

SOS_GAS_CONST
ANT_SEL

Selects use of pre-defined SOS Gas Constant or a
custom value entered in 'SOS Gas Constant'.

0 = 287 Jkg/K (Air), 1 = Custom

20959-01 Rev 02

Page 3-5