Table 5 Modbus Holding Registers (continued)

Address Size

Type Value

Description

Values

4071 2

Float

SOS_K_MIN

This value sets the lower limit in k-space that is used in
the SOS auto frequency determination code. This value
is equal to the first k-space bin after 0: PI/deltaX/50
(there are 50 bins from 0 to PI/deltaX).

4073 2

Float

SOS_K_MAX

This value sets the upper limit in k-space that is used in
the SOS auto frequency determination code. This value
is equal to the last k-space bin: PI/deltaX.

4075 2

Float

SOS_SEARCH_LIMI
T

This value is the +/- percentage of the estimated SOS
value (calculated using the auto frequency calculation
code) that determines the lower (Estimated SOS * 0.5)
and upper (Estimated SOS * 1.5) SOS search limits.

4077 2

Float

SOS_LAMBDA_DIA
METER

Used to calculate the SOS dynamic frequency maximum
used when calculating SOS. SOS Max Freq = (Max
SOS search) / ((Lambda Diameter * (Pipe Diameter/12)).

4079 2

Float

SENSOR_SPACING
_1

Starting point for sensor 1. Typically 0.

4081 2

Float

SENSOR_SPACING
_2

Distance in feet between sensor 1 and sensor 2.

4083 2

Float

SENSOR_SPACING
_3

Distance in feet between sensor 1 and sensor 3.

4085 2

Float

SENSOR_SPACING
_4

Distance in feet between sensor 1 and sensor 4.

4087 2

Float

SENSOR_SPACING
_5

Distance in feet between sensor 1 and sensor 5.

4089 2

Float

SENSOR_SPACING
_6

Distance in feet between sensor 1 and sensor 6.

4091 2

Float

SENSOR_SPACING
_7

Distance in feet between sensor 1 and sensor 7.

4093 2

Float

SENSOR_SPACING
_8

Distance in feet between sensor 1 and sensor 8.

4095 2

Float

SENSOR_SCALE_1

Scaling

factor in volts per PSI for sensor 1.

4097 2

Float

SENSOR_SCALE_2

Scaling

factor in volts per PSI for sensor 2.

4099 2

Float

SENSOR_SCALE_3

Scaling

factor in volts per PSI for sensor 3.

20959-01 Rev 02