TTL-level 0 V = CW rotation, TTL-level 5 V = CCW rotation.

-

A 5 volts TTL signal turns item ON, next signal to follow turns item OFF. This function is

available all the time (in internal mode either).

-

For triggering sense of rotation and ON / OFF function, a 5 volts pin is available in the

interface connector which allows potential-free switching.

-

The analog interface is recognized at a level of 0.2 volts or a 4 mAmps current; This mode is

confirmed by "EA" in the display. As long as the pump drive is controlled via the analog

interface,

"

Start/Stop

"

key will be the only active key on the item.

-

Wiring of the interface connector (37):

PIN 1 = +5 volts

PIN 2 = start/stop

PIN 3 = 0-10 V

PIN 4 = sense of rotation

PIN 5 = 0 V (GND)

PIN 6 = 4-20 mA

View on terminal end

-

Use IP 67 plug connectors only (Binder 6-pin 99-5121-15-06), Heidolph P/N 14-010-006-81,

not to void splash water protection.

-

Installation of plug connectors is limited to skilled personnel.

9.4.2. RS 232 interface (fig. 50)

-

The RS 232 interface allows pump control from a PC via parallel interface.

-

The RS 232 interface connector is installed in the rear panel of the pump drive. Wiring:

PIN 1 = PIN 1 9-pin.SUB-D

PIN 2 = PIN 6

PIN 3 = PIN 2

PIN 4 = PIN 7

PIN 5 = PIN 3

PIN 6 = PIN 8

PIN 7 = PIN 4

PIN 8 = PIN 5

View on terminal (soldered) end

-

Use IP 67 plug connectors only (Binder 8-pin 99-5171-15-08), Heidolph P/N 14-010-006-82,

not to void splash water protection.

-

Installation of plug connectors is limited to skilled personnel.

Interface parameters: 9600 Baud, no parity, 8 bit, 1 stop bit.

-

As long as the RS 232 interface is active, all of the pump's control elements and displays will

remain active.