Commands:

query

range

reply

remark

action

LEDx1?

x1=0...7

LED=0000\r\n

LED is „OFF"

LED-status

LED=0001\r\n

LED is „ON"

DSP?

DSP=xxxx

xxxx=display

act. displays


command

range

action

TAx2!

x2=0...7

key

0=no key 1=Cal

2=start 3=sense of rotation

4=max 5=up

6=down 7=select

SPx3=xxxx!

x3=0 xxxx=0001..5999 (0.1 sec)

pause

x3=2 xxxx=0010..0599 (min)

x4=3 xxxx=0010..0750 (hour)

SDM=xxxx!

xxxx=000,0..9999 (0.1 sec)

volume, metered

SDZ=xxxx!

xxxx=0000..9999 (depends on transmission used)

rotational speed

SMM=xxxx!

xxxx=000,1..9999 (not during calibration)

ml/min

wait a minute for reply.

SSD=xxxx!

xxxx=0000..0014

hose #

general

reply

remark

action

OK\r\n

command successful

self explaining

ERROR\r\n

command failed

none

PO\r\n

metering complete

none

9.5. Recommended sizes of tubes

-

To achieve high metering accuracy, in particular in the metering by volume mode, we

recommend the smallest tube diameter available; higher metering accuracy of small dia.

tubes is a function of the higher number of revolutions of the pump drive.

Advertising