Some or all transducer
output signals read 0 volts
at the TI output even when
a load is present.

The TI is not turned on.

Check that TI is turned on and the green power

indicator is lit.

The transducer signal cable
is not connected.

Check that the cable is connected between the
TI box and the slip ring (or connector housing

for a road simulator).

Output cables are not
connected.

Check that the output connectors are securely
fastened to the data acquisition device.

Cables or connectors are

damaged.

Check all cables and connectors, particularly the

power and transducer cable. Cables must be free
of nicks or cuts, and all connector pins shown

on the cable drawings must be present and not

bent.

Data acquisition

configuration, cabling, or

input ports are incorrect or
damaged.

Verify that the data acquisition is configured

correctly. While the TI is turned on and the

cable from the transducer to the TI is connected,
use a handheld DVM to verify output on

individual channels while they are being loaded.

Grounding problem.

Check that there is a good signal-to-ground
connection between the TI power cable and the

negative terminal on the battery

Troubleshooting Guide (part 2 of 10)