MTS SWIFT 10 ATV Sensor User Manual

Page 122

Advertising
background image

SWIFT 10 ATV Sensors

122

Troubleshooting

Output levels are much
higher or lower than
expected

Gains in the TI calibration

file have been overwritten

or modified.

Use the TI2XFER program to upload the

current TI calibration file (refer to

, “Upload the

Calibration File,”

on page 56). Check that the

gain settings match the original calibration file

sent with the transducer. If they are different,

download the original file to the TI, and make
any mode changes needed for your specific

application.

Do not make changes to the original
calibration file on the disk.

Changing the original file means you will lose the
original calibration data for your transducer.

Make a copy of the original file and make any
changes to the copy.

The data acquisition scales

are set incorrectly.

Check that the data acquisition scales are set

correctly. The full scale calibration range is
shown on the Calibration Report, with 10 V =

Full Scale (unless you requested custom

calibration output before calibration at MTS).

With the TI turned on and all cables from the
transducer to TI connected, use a handheld

DVM to check the output on individual

channels while they are being loaded to make
sure that the output at the TI box and data

acquisition correlate. (Use the BNC connectors
located at the back panel of the TI.)

The transducer was zeroed

with load applied (or a
different load than the

intended tare weight for

non-spinning applications
only).

Rezero the TI, being careful not to touch or load

the transducer during the zeroing procedure.

A zero offset is present in

the data.

See Zero Offset on the previous pages of this

Troubleshooting Guide.

Troubleshooting Guide (part 5 of 10)

S

YMPTOM

P

OSSIBLE

C

AUSES

S

OLUTION

CAUTION

Advertising