a) Select Station Setup button > Channels > Ch1 > Force > Channel Input Signals Button > Limits
tab.
b) Set the Upper Limit to .05 kN.
c) Set the Upper Limit Action to Interlock.
d) Set the Lower Limit to -.05 kN.
e) Set the Lower Limit Action to Interlock.
3. Set error detectors to 20% of half range to trip the interlock if the actuator goes unstable.
MTS TestLine Test System Operation | 59
Running the Example HCF Test