MTS TestLine User Manual

Page 94

Advertising
background image

Set up a Scope to compare the command signal to the resulting feedback. Enlarge the scope if
necessary, and then set up the scope as follows:

Y1 parameters:

Channel Ch1 (or other depending on your configuration)

Signal Command

Unit kN

Units/div 0.1000

Offset 0.3000

Y2 parameters:

Channel Ch1 (or other depending on your configuration)

Signal Force

Unit kN

Units/div 0.1000

Offset 0.3000

Time:

Trace Time 5.0000 s

5. Set the HPU and HSM to Power High.

94 | MTS TestLine Test System Operation

Running the Example HCF Test

Advertising