MTS TestLine User Manual

Page 99

Advertising
background image

Correct P Gain

Note: The scope display shows how well the feedback signal is following the command
signal. You should strive for a good match to the "Correct P Gain" example, realizing that
perfect adjustment is not possible.

8. When you are done adjusting the Force P Gain, adjust the Force I Gain to 25% of the P Gain value.

MTS TestLine Test System Operation | 99

Running the Example HCF Test