a) Open the Tuning: Ch1 Displacement window by selecting Station Setup button> Channels >

Ch1 (or other depending on your configuration) > Displacement, clicking the Tuning Fork icon,
and selecting the Adjustments tab.

b) Use the following parameters for this hard specimen test:

P Gain 10

I Gain 1

D Gain 0

F Gain 0

F2 Gain 0

Proportional FF Gain 0

FL Filter Set to maximum positive value

Note: The parameters shown are for the Example HCF Test using a hard specimen. For
other tests, we recommend initial settings based on known good settings from tests similar
to the one you are setting up to run. If you are unsure of what these initial settings should be,
contact MTS technical support.

3. Set Force tuning parameters.

MTS TestLine Test System Operation | 61

Running the Example HCF Test