MTS TestLine User Manual

Page 85

Advertising
background image

Because the specimen is fixed solidly in place, you can set narrow displacement limits that will trip if
anything breaks.

a) Click the Station Setup button.
b) Select Channels > Ch1 > Displacement > Channel Input Signals button > Limits tab.
c) Set the Upper Limit to 5 mm.
d) Set the Upper Action to Interlock.
e) Set the Lower Limit to -5 mm.
f) Set the Lower Action to Interlock.

2. Set force limits that are just outside the expected values for tuning.

MTS TestLine Test System Operation | 85

Running the Example HCF Test

Advertising