Page 14 – Practical Instrument Electronics 820-ELITE User Manual

Page 16

Advertising
background image

Page 14

READ

mA

18.203

SOURCE

HI

SET

LO

OFF

MAX
READ

MIN

READ

OVERLOAD

Hz SYNC

Double Click

Menu

Push & Turn

for

Fast Dialing
Push & Hold

to

Store/Step

mA ∙ V ∙ pH ∙ TC ∙ Ω ∙ RTD ∙ Freq ∙ Pressure

Milliamp Output Signal

Controller

Transmitter

P/I

DCS

READ mA, READ % (Percent of 4 to 20 mA)

Choose this function to measure from 0.000 to 24.000 milliamps
or -25.00 to 125.00%.

Move the power switch

w

to READ then Double Click the

EZ-DIAL knob to get into the Double Click Menu. Turn the knob

e

to scroll through the settings and press the knob to make your

selection. Select mA for the FUNCTION and READ for the
MODE. Choose either mA or % and whether you need the 250

HART resistor active in the loop.

Connect the red input lead (+) of the PIE 820-ELITE to the more
positive point of the break and the black input to the more
negative point.

Signals below 0 mA or open circuits are indicated by 0.00 mA
(-25.0%) on the display. Signals above 24 mA are current limited
by protection circuitry with “OVERRANGE” flashed on the
display and the red OVERLOAD LED lit.

The PIE 820-ELITE measures the input signal and constantly
updates the display with the current reading. Move the
EZ-CHECK switch

q

to MAX to see the highest reading and to

MIN to see the lowest reading. Press and hold the knob

e

to

clear the MAX and MIN readings.

All milliamp connections