Page 18, V & mv source – Practical Instrument Electronics 820-ELITE User Manual

Page 20

Advertising
background image

Page 18

Process Calibrator

SOURCE

HI

SET

LO

OFF

MAX
READ

MIN

READ

OVERLOAD

Hz SYNC

Double Click

Menu

Push & Turn

for

Fast Dialing
Push & Hold

to

Store/Step

READ

V

10.000

HI

mA ∙ V ∙ pH ∙ TC ∙ Ω ∙ RTD ∙ Freq ∙ Pressure

Voltage Receiver Input

Controller

Transmitter

Computer

Logger

DCS

V & mV Source

mV/V SOURCE

Choose this function to provide an output from -20.000 to 99.999
mV, -500.00 TO 999.99 mV or from 0.000 to 10.250 V. The
source current is a nominal 24 mA to provide the driving power
to your voltage receivers.

Move the power switch

w

to SOURCE then Double Click the

EZ-DIAL knob to get into the Double Click Menu. Turn the knob
to scroll through the settings and press the knob to make your
selection. Select V for the FUNCTION and 10V or 80 mV for the
RANGE.

Connect the output leads of the PIE 820-ELITE to the inputs of
the device being calibrated, making sure to check polarity. Red
lead to the plus (+) input and black lead to the minus (-) input.

Instantly output your SPAN and ZERO output settings by moving
the EZ-CHECK switch between HI and LO. You may also select
any third output setting (such as mid-range) using the SET
position on the EZ-CHECK switch. The output is adjusted in
0.001 mV, 0.01 mV or 0.001 V increments by turning the knob

e

.

Press and turn the knob for faster dialing with 0.100 mV, 1.00 mV
or 0.100 V increments.

Advertising