Thermocouple Source

Choose this function to provide a simulated thermocouple signal
into controllers, temperature transmitters, indicators or any input
devices that measure thermocouple sensors.

Move the power switch

w

to SOURCE then Double Click the

EZ-DIAL knob to get into the Double Click Menu. Turn the knob
to scroll through the settings and press the knob to make your
selection. Select T/C for the FUNCTION, °F or °C for the UNITS,
T/C Type (J, K, T, E, R, S, B, N, G, C, D, L (J-DIN), U (T-DIN) or
P (Platinel II) and internal COLD JUNC ON or OFF (ON is the
default).

Connect the PIE 820-ELITE to the inputs of the device being
calibrated using the proper type of thermocouple wire via the
miniature thermocouple socket.

Instantly output your SPAN and ZERO output settings by moving
the EZ-CHECK switch between HI and LO. You may also select
any third output setting (such as mid-range) using the SET
position on the EZ-CHECK switch. The output is adjusted in 0.1°
increments by turning the knob

e

. Press and turn the knob for

faster dialing with 10.0° increments.

All thermocouple connections

SOURCE

HI

SET

LO

OFF

MAX
READ

MIN

READ

OVERLOAD

Hz SYNC

Double Click

Menu

Push & Turn

for

Fast Dialing
Push & Hold

to

Store/Step

READ

TYPE K °C

820.0

LO

mA ∙ V ∙ pH ∙ TC ∙ Ω ∙ RTD ∙ Freq ∙ Pressure

Instrument with T/C Input

Controller

Temperature Transmitter

Temperature Indicator

Temperature Trip or Alarm

Thermocouple Wire

Advertising