Page 29

Read Frequency

Choose this function to count frequency.

Move the power switch

w

to READ then Double Click the

EZ-DIAL knob to get into the Double Click Menu. Turn the knob
to scroll through the settings and press the knob to make your
selection. Select FREQ for the FUNCTION and 20KHZ,
10000HZ, 1000HZ or 2000CPM for the RANGE.

Disconnect all input wires from the devices to be calibrated and
connect the PIE 820-ELITE to the output of the device matching
polarity.

The green HZ SYNC LED pulses in synch with the input
frequency.

Signals above the maximum scale are limited by protection
circuitry with “OVERRANGE” flashed on the display and the red
OVERLOAD LED lit.

The PIE 820-ELITE measures the input signal and constantly
updates the display with the current reading. Move the
EZ-CHECK switch

q

to MAX to see the highest reading and to

MIN to see the lowest reading. Press and hold the knob

e

to

clear the MAX and MIN readings.

All Frequency

Connections

SOURCE

HI

SET

LO

OFF

MAX

READ

MIN

READ

OVERLOAD

Hz SYNC

Double Click

Menu

Push & Turn

for

Fast Dialing
Push & Hold

to

Store/Step

READ

Hz

820.00

mA ∙ V ∙ pH ∙ TC ∙ Ω ∙ RTD ∙ Freq ∙ Pressure

Voltage Output Signal

Flowmeter

Flow Sensor

Variable Speed Drive

Controller

Transmitter

All Frequency

Connections