Page 31, For 2 seconds to clear the max and min readings – Practical Instrument Electronics 820-ELITE User Manual

Page 33

Advertising
background image

Page 31

Read Pressure
3) Connect pressure hoses, fittings & pumps (if required) to the
pressure instrument to be checked.

4) Move the switch on the Pressure Module to PSI and press the
E-Z Dial knob on the 820-ELITE.

5) Press and hold the

e

E-Z DIAL KNOB for 1 second to 'Zero'

the pressure. The word "TARE" appears on the display indicating
that all measurements are relative to the pressure measured
when the calibrator was zeroed.

Pressure readings appear on the display of the 820-ELITE and
are constantly updated.

The PIE 820-ELITE measures the pressure and constantly
updates the display with the current reading. Move the
EZ-CHECK switch

q

to MAX to see the highest reading and to

MIN to see the lowest reading. Press and hold the knob

e

for 2

seconds to clear the MAX and MIN readings.

30 psi/2 bar Gauge &

Absolute Modules

Units

Maximum

PSI

30.000

Inches H2O@4°C

830.40

Inches H2O@20°C

831.90

Inches H2O@60°F

831.22

Inches HG@0°C

61.080

bar

2.0684

mbar

2068.4

kPa

206.84

MPa

0.2068

Kgf/CM2

2.1092

cm H2O@4°C

2109.2

cm H2O@23°C

2114.4

mm H2O@4°C

21092

mm H2O@23°C

21144

mm Hg @ 0°C

1551.4

500 psi/35 bar Gauge Modules

Units

Maximum

PSI

500.00

Inches H2O@4°C

999.99*

Inches H2O@20°C

999.99*

Inches H2O@60°F

999.99*

Inches HG@0°C

1018.0

bar

34.474

mbar

34474

kPa

3447.4

MPa

3.4474

Kgf/CM2

35.153

cm H2O@4°C

35153

cm H2O@23°C

35240

mm H2O@4°C

99999*

mm H2O@23°C

99999*

mm Hg @ 0°C

25857

*

Indicates OVER RANGE above this value